Teachnical Data
Dimensions
HC2A-IE02-NPT: NPT thread x 104 mm x 2 m Cable
HC2A-IE02-G : G thread x 104 mm x 2 m Cable
Probe type
Industrial probe
Humidity sensor
Hygromer HT-1
Temperature sensor
PT100 1/3 DIN Class B
Long-term stability
<1 %rh / year
Measuring range
-100…200 °C (short-term peak load) / 0…100 %rh
Accuracy w. adjustment profile “standard”
at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K
Psychrometric calculations
Dew or frost point
Analogue outputs scaleable by the user
Yes
Software Update
With internet and HW4 software
User information
from Software HW4 (devicemanager)
FDA/GAMP compatibility
FDA 21 CFR Part 11 and GAMP compatible
Probe adjustment by software
1 Point & multipoint %rh & °C
Analog output signal (standard)
0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Reverse voltage protection
Mechanically in the connector
Max. cable length
Up to 5 m with passive cable. Up to 100 m with active cable.
Probe material
PPS, Stainless steel 1.4301
Probe protection
Filter carrier, brass nickel plated
Filter cartridge
depending on the application, order separately
Power supply
3.3…5 VDC
Pressure range
0 – 100 bar