Specifications
Model
| 729 30G FC |
|
||||||||
| 729 150G FC |
|
||||||||
| 729 300G FC |
|
| Electrical specification | |
| All specifications are valid to 110% of range, except 24 mA source and simulate which are valid to 100% of range. | |
| Ranges | |
| mA measure, source and simulate | 0 mA to 24 mA |
| Volts dc measure | 0 V dc to 30 V dc |
| Resolution | |
| mA dc source, simulate and measure | 1 μA |
| Voltage dc measurement | 1 mV |
| Accuracy | 0.01% ±2 LSD all ranges (at 23 °C ±5 °C) |
| Stability | 20 ppm of full scale /°C from -10 °C to +18 °C and 28 °C to +50 °C |
| mA simulate external voltage requirement | 12 V dc to 30 V dc |
| Loop compliance voltage | 24 V dc @ 20 mA |
| mA simulate external voltage requirement | 12 V dc to 30 V dc |
| Temperature measurement only/100 Ω Pt(385) RTD | -50 °C to 150 °C (-58 °F to 302 °F) |
| Temperature resolution | 0.01 °C (0.01 °F) |
| Temperature accuracy | ±0.1 °C (0.2 °F) ±0.25 °C (±0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory) |
| Drive capability | 1200 Ω without HART resistor, 950 Ω with internal HART resistor |
| Mechanical specification | |
| Size (HxWxL) | 7 x 27.94 x 17.27 cm (2.75 x 11 x 6.8 in) |
| Weight | 6.5 lbs., 2.95 kg |
| Ingress protection | IP54 |
| Environmental specification | |
| Operating temperature | -10 °C to 50 °C for measurement, 0 °C to 50 °C for pressure control |
| Operating temperature with battery | -10 °C to 40 °C |
| Battery will only charge to 40 °C | |
| Storage temperature | -40 °C to 60 °C |
| Operating altitude | 3000 m |
| Storage altitude | 13000 m |
| Wireless (729 FC only) | |
| Radio frequency certification | FCC ID: T68-FBLE IC:6627A-FBLE |
| Wireless radio frequency range | 2412 MHz to 2462 MHz |