Specifications
[1] Control Stability is based on the range of pressure control module
Pressure Specifications
Pressure Range
15-10,000 psi (1-700 Bar)
Range Selection
Manual 3K psi, Manual 10K psi or Auto-range
Accuracy
0-10,000 psi, 0.01%FS
0-10,000 psi, 0.02%FS
0-3,000 psi, 0.01%FS
0-3,000 psi, 0.02%FS
Resolution
7 digits
6 digits
Maximum External Load Capacity
Max: 80 ml @ 700 Bar, 50 ml reccomended
Reservoir
Max: 350 ml, push-on reservoir cap, built-in filter
Control Stability [1]
≤ 0.005% FS from 100 to 10,000 psi
Stability Duration
> 5 min
Pressure Modules
Built-in one module with dual range
External Control Pressure Module
See the following “External Control Pressure Module Specification and Compatibility” table
External Measurement Pressure Module
All ADT161 pressure modules
Electrical Specifications
Specification
Range
Resolution
Accuracy
Note
Current Measure
Auto-ranging ±25 or ±50 mA
0.1 μA
0.008 %RD + 1.0 μA
Impedance < 10 Ω
0.008 %RD + 2.0 μA
Voltage Measure
±300 mV
1 μV
0.008 %RD + 6 μV
Impedance > 1 GΩ
Auto-ranging ±5, ±12 or ±30 V
±5 V: 20 μV
0.008 %RD + 100 μV
Impedance > 1 MΩ
±12 V: 0.1 mV
0.008 %RD + 240 μV
±30 V: 0.1 mV
0.008 %RD + 600 μV
Loop Power Source
24 V
N/A
±1 V
50 mA (Max Loading)
mA Source
0 to 2.5 mA or 2.5 to 25 mA
0-2.5 mA: 0.05μA, 0-25 mA: 0.5μA
0-2.5 mA: 0.008 %RD + 0.2 μA, 0-25 mA: 0.008 %RD + 1.0 μA
20 mA @ 1 kΩ
Power Source
16 to 30 V
1 V
±1 V
70 mA (Max Loading)
V Source
0 to 16 V
250 μV
0.008 %RD + 500 μV
Pressure Switch
Mechanical Switch, Live Mechanical Switch, NPN Switch, PNP Switch
N/A
N/A
Response time < 10 ms if the switch is powered. Voltage range: (3-30) V
Temperature Compensation
18°C to 28°C
Temperature Coefficient
Outside of 18°C to 28°C: <±0.0005%RD+0.00005%FS/°C
Misuse Protection
Up to 30 V on any two sockets
Pressure Switch Test
Supported
HART / Profibus PA
Supported
General Specifications
Specification
Description
User Interface
Color touch screen and keypad operation
Display
7” TFT touch screen 800 x 480 color
Enclosure IP Rating
IP31
Power
Dedicated lithium battery or power adapter
Battery
Rechargeable Li-Ion battery, typically 12 hours of operation, less than 5 hours recharge.
Weight
28 lbs (12.8 kg) without media
Size
11.81 x 8.66 x 7.56 in (300 x 220 x 192 mm)
Communications
Wi-Fi, LAN, Bluetooth, USB, Ethernet
HART Communication
Read, configure and calibrate HART devices – DD files updated periodically
Data Storage
> 8 GB
Data Logging
Up to 1,000,000 readings (data and time stamped)
Task documentation
Up to 1,000 tasks can be stored with data
Automation Functions
Switch test, auto step, leak test
User Interface Localization
English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian
Pump life
> 1,000,000 cycles
Environmental Specifications
Operating temperature: 32°F to 122°F( 0°C to 50°C)
Storage temperature: -20°C to 60°C ( -4°F to 120°F)
Humidity: <90%, non-condensing
Certification
ISO 17025 accredited certificate of calibration with NIST-traceable data
Compliance
CE
Software Compatibility
ACal, Additel Land and Additel Link for access via mobile application